Table 7. Bias (calculated as the difference between the criterion ABL90 and each analyzer) for blood lactate concentration, and the effect of the bias on power at calculated Lactate Threshold One (LT1) and Two (LT2). Bias calculations used the average of Device A and B combined. Threshold calculations employed the ‘modified’ Dmax method in ADAPT software (Australian Institute of Sport, 1995) using representative cycling and rowing data from incremental step tests.
Lactate concentration band (mM) Analyser
ABL90 Edge i-STAT Lactate Pro Lactate Pro2 Scout+ Xpress
Bias (mM)
0-1.9 0.0 0.34 -0.12 -0.17 0.32 -0.17 0.06
2.0-4.9 0.0 0.07 -0.29 -0.70 -0.06 -0.89 -0.13
5.0-9.9 0.0 -0.32 0.17 -1.76 -1.48 -2.15 -0.09
10.0-14.9 0.0 -0.29 0.00 -2.41 -1.81 -2.71 -0.70
15.0+ 0.0 -2.05 0.62 -5.40 -2.16 -6.15 -6.70
Threshold intensity Cycling
LT1 Power (W) 189 191 194 191 187 192 190
LT2 Power (W) 249 249 253 249 249 249 226
Rowing
LT1 Power (W) 206 209 209 211 207 215 209
LT2 Power (W) 268 277 263 281 281 286 263