Research article - (2008)07, 15 - 22
Lack of Agreement Between Gas Exchange Variables Measured by Two Metabolic Systems
Djordje G. Jakovljevic1,, David Nunan1, Gay Donovan1, Lynette D. Hodges2, Gavin R. H. Sandercock3, David A. Brodie1
1Buckinghamshire New University, Buckinghamshire, UK
2University of Bedfordshire, Bedfordshire, UK
3University of Essex, Colchester, UK

Djordje G. Jakovljevic
✉ Research Centre for Society and Health, Buckinghamshire New University, Gorelands Lane, Chalfont St Giles, Buckinghamshire HP8 4AD, UK.
Email: djordje.jakovljevic@bcuc.ac.uk
Received: 10-09-2007 -- Accepted: 30-10-2007
Published (online): 01-03-2008

ABSTRACT

The purpose of this study was to assess the agreement and consistency between gas exchange variables measured by two online metabolic systems during an incremental exercise test. After obtaining local ethics approval and informed consent, 15 healthy subjects performed an incremental exercise test to volitional fatigue using the Bruce protocol. The Innocor (Innovision, Denmark) and CardiO2 (Medical Graphics, USA) systems were placed in series, with the Innocor mouthpiece attached to the pneumotach of the CardiO2. Metabolic data were analysed during the last 30 seconds of each stage and at peak exercise. There were non- significant differences (p > 0.05) between the two systems in estimation of oxygen consumption (VO2) and in minute ventilation (VE). Mean Cronbach’s alpha for VO2 and VE were 0.88 and 0.92. The Bland-Altman analysis revealed that limits of agreement were -0.52 to 0.55 l.min-1 for VO2, and -8.74 to 10.66 l.min-1 for VE. Carbon dioxide production (VCO2) and consequently respiratory exchange ratio (RER) measured by the Innocor were significantly lower (p < 0.05) through all stages. The CardiO2 measured fraction of expired carbon dioxide (FeCO2) significantly higher (p < 0.05). The limits of agreement for VO2 and VE are wide and unacceptable in cardio-pulmonary exercise testing. The Innocor reported VCO2 systematically lower. Therefore the Innocor and CardiO2 metabolic systems cannot be used interchangeably without affecting the diagnosis of an individual patient. Results from the present study support previous suggestion that considerable care is needed when comparing metabolic data obtained from different automated metabolic systems.

Key words: Metabolic system, oxygen consumption, minute ventilation, carbon dioxide production, Bruce protocol

Key Points
  • There is general concern regarding the limited knowledge available about the accuracy of a number of commercially available systems.
  • Demonstrated limits of agreement between key gas exchange variables (oxygen consumption and minute ventilation) as measured by the two metabolic systems were wide and unacceptable in cardio-pulmonary exercise testing.
  • Considerable care is needed when comparing metabolic data obtained from different automated metabolic systems.