Intramuscular Temperature Differences Between the Mid-Point and Peripheral Effective Radiating Area With Ultrasound
Michale G. Miller, Janae R. Longoria, Christopher C. Cheatham, Robert J. Baker, Timothy J. Michael
Department of HPER, Western Michigan University, Kalamazoo, MI, USA

Michale G. Miller
✉ Associate Professor, Western Michigan University, Dept of HPER, 1903 W Michigan Ave, Kalamazoo, MI 49008-5426, USA
Email: michael.g.miller@wmich.edu
Received: 01-04-2008 -- Accepted: 24-04-2008
Published (online): 01-06-2008

ABSTRACT

The purpose of the study was to determine whether uniform intramuscular heating is achieved throughout a treatment area 2 times the transducer head at both 1 and 3 MHz. Seven male and three female subjects (Age: 23.6 ± 1.0 yrs, Weight: 83.8 ± 23.2 kg, Site Skinfold: 13.9 ± 7.3 mm) underwent two ultrasound treatments (1 and 3 MHz) in the triceps surae muscle group. Thermocouples were inserted at the midpoint and periphery of the treatment area. Ten minute baseline temperatures were recorded followed by a ten minute ultrasound treatment. Two (site) X 10 (time) repeated measures ANOVAs were separately used to determine significance for 1 and 3 MHz treatments. Post-hoc testing was performed using the Bonferroni adjustment. A significant site-by-time interaction was observed for both the 1 and 3 MHz treatments. From baseline to the end of the treatment, temperature increased approximately 2.62°C and 1.58°C for the midpoint and periphery of the 1 MHz treatment and 5.88°C and 3.64°C for the 3 MHz treatment. The differences in temperature suggest that uniform heating does not occur throughout the treatment area.

Key words: Tissue, thermal effects, acoustical.

Key Points
  • 3 MHz is more effective in raising intramuscular temperature within ERA.
  • Stroke count/rate of transducer may play a factor in heating tissue.
  • Treatment size may alter uniform heating.
Subjects

Ten subjects, seven male and three females (23.6 ± 1.0 years, height: 1.74 ± 0.09 m, weight: 83.8 ± 23.2 kg, skinfold thickness: 13.9 ± 7.3 mm, calf circumference: 37.7 ± 4.5 cm) volunteered for the study. Subjects were verbally screened for prior history of trauma to the leg, recent leg injuries within the past 6 months, infection, and vascular or nervous conditions. The number of subjects was based on a power analysis completed prior to recruitment using the means and standard deviations of previous research (Ashton et al., 1998; Bishop et al., 2004; Burr et al., 2004; Chan et al., 1998; Draper et al., 1995; Myrer et al., 2001; Oshikoya et al., 2000; Rose et al., 1996). The procedures of this study were approved through the Human Subjects Institutional Review Board and informed consent was obtained prior to subject participation.

Instruments

An Omnisound 3000 (Accelerated Care Plus, Reno, NV) ultrasound unit with a beam non-uniformity ratio of 2:1 for 1 MHz and 3:1 for 3 MHz was used for both frequencies. Using the 5-cm2 sound head, a rectangular template was made to measure an area of two times the sound head, or approximately 7.2 cm x 3.5 cm. For the coupling medium, we utilized Aquasonic coupling gel (Parker Laboratories, INC, Fairfield, NJ) at a consistent temperature of 22.4°C ± 0.43°C for the 1 MHz and 3 MHz treatments. Ambient temperature was recorded using a portable thermometer (Omega Microprocessor Thermometer, Stamford, CT). Intramuscular temperatures were measured using two, 26-gauge, 4 cm long thermocouple needles (MT-26/4 hypodermic needle microprobe, Physitemp, Clifton, NJ) connected to an electrothermometer (Iso-Thermex, Columbus Instruments, Columbus, OH) that recorded temperatures at 1 minute intervals. We utilized a template that fit around the calf muscle and measured depth in centimeters to determine the point of entry for each thermocouple needle (Figure 1).

Procedures

Our treatment table was set up in a room with an ambient temperature of 22.3°C ± 0.4°C. With the subjects lying prone, we determined the widest part of the calf muscle and placed the treatment template on the surface of the calf. We found the middle of the template then placed the depth template to mark the insertion for the first thermocouple needle (2.5 cm for the 1-MHz treatment, and 1.0 cm for the 3-MHz treatment) to stay within depths commonly used as guides when performing ultrasound treatments (Prentice, 2003). We aligned the depth template along the distal edge of the rectangular template and marked the second insertion point based upon the depth targeted. The template was aligned so that when the probes were inserted to the maximal length, it fell within the middle of the template. The probes were placed approximately 3.6 cm apart within the template area (Figure 2).

Both the treatment area and the thermocouple insertion marks were shaved, cleansed with a 0.40% benzalkonium chloride antiseptic towelette, and re-cleansed with a 70% isopropyl alcohol wipe. We inserted the entire 4 cm length of the thermocouple needles, one at the midpoint and the other at the periphery or distal end of the treatment site. Once connected to the electrothermometer, a 10-minute baseline period began and temperatures were recorded. The patients were in this position for study preparation and baseline measurements for approximately 30 minutes prior to the beginning of data collection. The ultrasound treatment parameters were set according to previous literature, 1 MHz at 1.5W/cm2, continuous, for 10 minutes (Ashton et al., 1998; Draper et al., 1995; Rose et al., 1996) and 3 MHz at 1.0W/cm2, continuous, for 10 minutes (Bishop et al., 2004; Draper et al., 1995; Leung et al., 2006). A one minute transition period between the baseline and the treatment was used to apply the coupling gel. Movement of the transducer head was timed to 4 cm/sec (Draper, 1998; Weaver et al., 2006), using a metronome (Yamaha QT-1 quartz metronome) in longitudinal patterns across the template, ensuring that the edges of the treatment area were covered. Intramuscular temperatures were recorded every minute from the beginning of baseline to the end of the treatment time. All subjects completed the 1 MHz ultrasound treatment first, followed by the 3 MHz treatment with a minimum of 48 hours between treatments. The subjects were not counterbalanced since the 3 MHz treatment was HSIRB approved and initiated several days after the original study of just measuring 1 MHz differences.

At the end of the treatment, the coupling gel and thermocouple needles were removed and the area re-cleansed with a 70% isopropyl alcohol wipe then soaked and cleaned with Cidex OPA (Johnson & Johnson, Irvine, CA) as recommended by the manufacturer. Standard adhesive medical strips were applied to the thermocouple insertion points and subjects were instructed on the use of ice to relieve muscle soreness if occurred.

Statistical analysis

We used a 2X10 ANOVA with repeated measures to determine differences in intramuscular temperature between the two measurement sites (midpoint and periphery) and time (minutes 1 through 10) for both 1 MHz and 3 MHz ultrasound treatments separately. All statistical testing was conducted using the Statistical Package for Social Sciences (SPSS version 14.0, Chicago, IL). The alpha level was set a priori at p < 0.05 to minimize type I errors. In the case of significant interactions, post-hoc testing was performed using a simple effects analysis with the Bonferroni adjustment at p = 0.0045. All data is presented as mean ± standard deviation units.

INTRODUCTION

Ultrasound is both a non-thermal and thermal modality which utilizes acoustic energy to promote healing (Prentice, 2003). Its non-thermal effects include collagen alignment (Byl et al., 1993; da Cunha et al., 2001), increased collagen and wound strength (Byl et al., 1993) acoustic streaming (Garrett et al., 2000; Kramer, 1984), cavitation (Prentice, 2003; Johns, 2002), decreased nerve conduction velocity (Kramer, 1984), peripheral nerve regeneration (Raso et al., 2005), bone fracture repair (Stein and Lerner, 2005; Warden, 2003), accelerating ligament healing (Warden et al., 2006), and edema reduction (Prentice, 2003). The thermal benefits of ultrasound include the treatment of muscle pain (Almeida et al., 2003), increased elasticity of both muscle and tendon (Chan et al., 1998; Draper et al., 1998; Draper and Ricard, 1995; Rose et al., 1996), increased tensile strength and energy absorption in tendons (Enwemeka, 1989), increased passive range of motion in joints (Knight et al., 2001), increased nerve conduction velocity (Kramer, 1984), increased ligament repair time (Leung et al., 2006), increased blood flow (Noble et al., 2007), and increased active range of motion in joints (Oates and Draper, 2006).

Thermal effects are often used to increase range of motion (Draper et al., 1998) and extensibility of collagen tissue (Chan et al., 1998; Draper and Ricard, 1995; Rose et al., 1996), therefore, it is imperative that the treatment area reach the critical temperature value in order to produce the desired physiological effects to lengthen these tissues. Previous research has suggested that a 3-4°C increase from baseline temperature is needed to increase tissue extensibility (Castel, 1993; Draper et al., 1995; Oates and Draper, 2006). If the goal is to increase the extensibility of tissue, it is crucial to heat the entire treatment area to cover as much connective tissue as possible to produce these desired effects. To our knowledge, only three studies (Fincher et al., 2007; Garrett et al., 2000; McCutchan et al., 2007) examined the differences in site (middle and proximal/distal) and found no temperature differences. However, two (Fincher et al., 2007; McCutchan et al., 2007), used an Autosound™ with different intervention methods and one measured the temperature distances over 40 times the effective radiating area (ERA) (Garrett et al., 2000).

Most clinicians apply ultrasound within their specific treatment area, often moving the transducer head in circular or longitudinal strokes in the middle of the treatment site more frequently than the outer edges. This pattern of movement will theoretically result in more transducer strokes in the middle of the treatment area versus the peripheral sites, affecting the temperature of the entire treatment area. Therefore, the purpose of our study was to determine if there is uniform intramuscular heating between the midpoint and periphery of a treatment area 2 times the size of 5cm2 traducer head during ultrasound treatments. Uniform heating is defined as the entire treatment site having the same temperature. We hypothesized that the midpoint of the 1 and 3 MHz treatment areas will have higher intramuscular temperatures compared to the periphery due to the higher occurrence of the transducer head moving across the middle of the treatment site.

METHODS
RESULTS

The responses in temperature within the middle and peripheral treatment sites during both 1 and 3 MHz frequency ultrasound treatments are illustrated in Figures 3 and 4">4, respectively. The two way repeated measures ANOVA for the 1 MHz treatment revealed a site by time interaction, (F10,90 = 25.66, p < 0.001, partial eta squared = 0.740). Post-hoc testing revealed that the temperature at the midpoint was greater than the periphery at all time points except for baseline and minutes 1 through 3 of the treatment. From baseline to the end of the treatment (10 min), temperature increased approximately 2.62°C at the midpoint and 1.58°C for the periphery.

The two way repeated measures ANOVA for the 3 MHz treatment revealed a site by time interaction, (F10,90 = 16.66, p<0.001, partial eta squared = 0.649). Post-hoc testing revealed that the temperature at the midpoint was greater than the periphery at all time points except for baseline and minute 1 of the treatment. From baseline to the end of the treatment (10 min), temperature increased at the midpoint approximately 5.88°C and 3.64°C for the periphery.

DISCUSSION

Our midpoint temperature increased approximately 2.62°C above the baseline in the middle of the treatment area with a rate of 0.262°C per minute for the 1 MHz treatment which is lower than results of previous research (Burr et al., 2004; Draper et al., 1995; Weaver et al., 2006). This may be explained by the variability of tissue types of the subjects. Our thermocouple placements may not have been consistently placed within the muscle tissue at both the midpoint and peripheral areas at the same depth because we only measured skinfold at the middle of the treatment area. It may be prudent to measure skinfold sites at both the midpoint and periphery sites to ensure that depth placement is consistent as the calf becomes narrower toward the distal treatment sites. In addition, at the periphery, we had a total increase from baseline only 1.58°C at a rate of 0.158°C per minute, suggesting that the periphery of the treatment site for 1 MHz did not reach the desired therapeutic effect and ranges as noted by previous research for tissue extensibility (Castel, 1993; Draper et al., 1995; Oates et al., 2006). However, our heating rates were based upon total time divided by total temperature increase and recent evidence suggest (Demchak et al., 2007) that heating rates per unit of time are curvilinear, not linear. More time may be needed for the periphery to reach the critical level and thus change the length of time ultrasound treatments are required for the treatment site to be uniformly heated.

For our 3 MHz treatment, at 1.0 W/cm2 for 10 minutes, we achieved an averaged increase of 5.88°C at a rate of 0.588°C per minute at the midpoint and an averaged increase of 3.64°C with a rate of 0.364°C per minute at the periphery. Our temperature increases supports current literature (Draper et al., 1995; Draper and Ricard, 1995), which states that intramuscular temperatures can be raised an average of 5.3°C using 3 MHz for ten minutes. However, it differs with the heating rate on tendons at a rate of 2.1°C/min (Chan et al., 1998). Our peripheral probes were placed distally to the midpoint, closer to the muscultotendinous junction of the triceps surae muscle group that may have accounted for temperature differences as a result of change in tissue density from muscle to tendon. Of importance is that although the periphery was 2.24°C below the mid point at the end of the 3 MHz treatment, it was still within the desired therapeutic range for vigorous heating (3-4°C) in order to facilitate tissue extensibility.

We used a treatment site 2 times the transducer head for both ultrasound treatments because most clinicians and researchers have commonly used this method of determining the treatment area (Holcomb, 2003; Holcomb and Joyce, 2003; Merrick et al., 2003; Noble et al., 2007), and it closely approximates a 2-3 ERA site used in previous studies (Ashton et al., 1998; Bishop et al., 2004; Byl et al., 1993; Draper et al., 1995; Gallo et al., 2004; Rose et al., 1996). Knowing that the ERA is somewhat less than the treatment area and that we did not have access to a digital power meter to specifically measure the ERA, we made sure to cross the edges of the treatment site by approximately 1-3 millimeters. These factors may explain the differences in our midpoint and peripheral temperatures. Our peripheral edges may not have received all the ultrasound energy and resulted in less temperature increases compared to the mid point where we were assured that the ultrasound energy was delivered. McCutchan et al., 2007 found no difference between sites, however, their thermocouples were placed approximately 2.9 cm apart which may have been too close to the midpoint to detect a difference. They also used an Autosound™ unit which does not utilize stroke/transducer head movement. Future studies should also record temperatures from the midpoint of the ERA to the periphery segmentally by centimeters to help determine exactly where in the treatment site temperature uniformity subsides.

The difference in intramuscular temperatures found between midpoint and periphery in our study may also be linked to the stroke count across the mid point versus the periphery or transducer velocity over the treatment area. Since the midpoint of the treatment site received more stroke counts than the periphery, it is understandable that the midpoint had higher temperatures. We used a metronome to provide a constant rhythm for us to follow over the treatment area at a rate of 4cm/s as recommended by previous research to provide even distribution of sound waves to control for transducer velocity (Bishop et al., 2004; Demchak et al., 2007; Draper, 1998; Draper et al., 1998). However, recent evidence suggests that transducer velocity does not alter intramuscular heating temperatures (Weaver et al., 2006). By experimenting with velocities of 2-3cm/s, 4-5cm/s, and 7-8cm/s, the researchers found that intramuscular temperatures after each change in velocity were within 1°C of each other based upon midpoint measurements. We used longitudinal strokes in our study but many clinicians use either longitudinal, circular, or a combination of both when performing ultrasound treatments. The specific type of stoke used for applying ultrasound should be investigated further.

The intramuscular temperature differences may also be a component of the specific type of ultrasound unit used. In accordance with the majority of the literature (Byl et al., 1993; Draper et al., 1995; Gallo et al., 2004; Garrett et al., 2000; Holcomb and Joyce, 2003; Oshikoya et al., 2000; Weaver et al., 2006), we used the Omnisound 3000 to maximize energy delivery. While the Omnisound 3000 has been shown to heat more efficiently than the Forte 400 Combo (Holcomb and Joyce, 2003), the heating effects delivered by other similar ultrasound units may vary (Johns et al., 2007a; 2007b), and further investigation into the differences among types of units as they relate to uniform heating throughout the ERA should be conducted.

CONCLUSION

When the goal is to improve extensibility via vigorous heating, it is important that the temperature is elevated by 3- 4°C throughout the treatment area. We found that 3 MHz, using standard parameters effectively achieved the desired therapeutic range at the midpoint and the periphery, suggesting that all the tissues in the treatment area met the 3-4°C critical value and would respond accordingly to stretching or elongation as desired. However, when using 1 MHz, we found that the periphery did not reach the critical temperature ranges. These results suggest that using ultrasound for heating deeper tissues within the desired therapeutic range of 3-4°C did not occur for 1 MHz and that all the tissues in the treatment area may not respond therapeutically as desired. If the goal is to produce deep heating in tissues uniformly, clinicians may need to select other modalities that can achieve these effects.

ACKNOWLEDGEMENTS

This study was partially funded by the Graduate Student Research Grant at the University.

AUTHOR BIOGRAPHY
     
 
Michale G. Miller
 
Employment:Associate Professor, WMU, Dept of HPER.
 
Degree: EdD, ATC, CSCS
 
Research interests: Aquatic plyometrics, electrical modalities, strength and conditioning.
  E-mail: michael.g.miller@wmich.edu
   
   

     
 
Janae R. Longoria
 
Employment:WMU, Dept of HPER.
 
Degree: MA, ATC
 
Research interests: Modalities and clinical education.
  E-mail: janae.longoria@wmich.edu
   
   

     
 
Christopher C. Cheatham
 
Employment:Associate Professor of Exercise Science, WMU, Dept of HPER.
 
Degree: PhD
 
Research interests: Environmental physiology, nicotine and human performance, influence of exercise on plasma and blood volumes.
  E-mail: chris.cheatham@wmich.edu
   
   

     
 
Robert J. Baker
 
Employment:Program Director Primary Care Sports Medicine Fellowship, Department of Family Medicine, MSU-Kalamazoo Center for Medical Studies. Team Physician Western Michigan University. Asssociate Professor of Clinical Medicine, Michigan State University.
 
Degree: MD, PhD, ATC, FACSM, FAAFP
 
Research interests: Primary care sports medicine, exercise physiology, modalities.
  E-mail: Baker@kcms.msu.edu
   
   

     
 
Timothy J. Michael
 
Employment:Associate Professor, WMU, Dept of HPER.
 
Degree: PhD
 
Research interests: Sport performance and nutrition, pediatric exercise science, clinical aspects of exercise, perception of exertion.
  E-mail: tim.michael@wmich.edu
   
   

REFERENCES
Almeida T.F., Roizenblatt S., Benedito-Silva A.A., Tufik S. (2003) The effect of combined therapy (ultrasound and interferential current) on pain and sleep fibromyalgia. Pain 104, 665-672.
Ashton D.F., Draper D.O., Myrer J.W. (1998) Temperature rise in human muscle during ultrasound treatments using flex-all as a coupling agent. Journal of Athletic Training 33, 136-140.
Bishop S., Draper D.O., Knight K.L., Feland J.B. (2004) Human tissue-temperature rise during ultrasound treatments with the aquaflex gel pad. Journal of Athletic Training 39, 126-131.
Burr P.O., Demchak T.J., Cordova M.L., Ingersoll C.D., Stone M.B. (2004) Effects of altering intensity during 1-MHz ultrasound treatment on increasing triceps surae temperature. Journal of Sport Rehabilitation 13, 275-286.
Byl N.N., McKenzie A., Wong T., West J., Hunt T.K. (1993) Incisional wound healing: A controlled study of low and high dose ultrasound. Journal of Orthopaedic Sports Physical Therapy 18, 619-628.
Castel J.C. (1993) Therapeutic ultrasound. Rehabilitation Therapy Product Review , 22-32.
Chan A.K., Myrer J.W., Measom G.J., Draper D.O. (1998) Temperature changes in human patellar tendon in response to therapeutic ultrasound. Journal of Athletic Training 33, 130-135.
da Cunha A., Parizotto N.A., Vidal B.C. (2001) The effect of therapeutic ultrasound on repair of the Achilles tendon (tendo calcaneus) of the rat. Ultrasound in Medicine & Biology 27, 1691-1696.
Demchak T.J., Straub S.J., Johns L.D. (2007) Ultrasound heating is curvilinear in nature and varies between transducers from the same manufacturer. Journal of Sport Rehabilitation 16, 122-130.
Draper D.O. (1998) Guidelines to enhance therapeutic ultrasound treatment outcomes. Athletic Therapy Today 3, 7-11.
Draper D.O., Anderson C., Schulthies S.S., Ricard M.D. (1998) Immediate and residual changes in dorsiflexion range of motion using an ultrasound heat and stretch routine. Journal of Athletic Training 33, 141-144.
Draper D.O., Castel J.C., Castel D. (1995) Rate of temperature increase in human muscle during 1MHz and 3MHz continuous ultrasound. Journal of Orthopaedic Sports Physical Therapy 22, 142-149.
Draper D.O., Ricard M.D. (1995) Rate of temperature decay in human muscle following 3MHz ultrasound: The stretching window revealed. Journal of Athletic Training 30, 304-307.
Draper D.O., Schulthies S., Sorvisto P., Hautala A. (1995) Temperature changes in deep muscles of humans during ice and ultrasound therapies: An in vivo study. Journal of Orthopaedic Sports Physical Therapy 21, 153-157.
Enwemeka C.S. (1989) The effects of therapeutic ultrasound on tendon healing. American Journal of Physical Medicine Rehabilitation 68, 283-287.
Fincher A.L., Trowbridge C.A., Ricard M.D. (2007) A comparison of intramuscular temperature increases and uniformity of heating produced by hands-free Autosound™ and manual therapeutic ultrasound techniques. Journal of Athletic Training 42, S41-.
Gallo J.A., Draper D.O., Brody L.T., Fellingham G.W. (2004) A comparison of human muscle temperature increases during 3-MHz continuous and pulsed ultrasound with equivalent temporal average intensities. Journal of Orthopaedic Sports Physical Therapy 34, 395-401.
Garrett C.L., Draper D.O., Knight K.L. (2000) Heat distribution in the lower leg from pulsed short-wave diathermy and ultrasound treatments. Journal of Athletic Training 35, 50-55.
Holcomb W.R. (2003) The effects of superficial heating before 1-MHz ultrasound on tissue temperature. Journal of Sport Rehabilitation 2, 95-103.
Holcomb W.R., Joyce C.J. (2003) A comparison of temperature increases produced by 2 commonly used ultrasound units. Journal of Athletic Training 38, 24-27.
Johns L.D., Straub S.J., Howard S.M. (2007a) Analysis of effective radiating area, power, intensity, and field characteristics of ultrasound transducers. Archives of Physical Medicine and Rehabilitation 88, 124-129.
Johns L.D., Straub S.J., Howard S.M. (2007b) Variability in effective radiating area and output power of new ultrasound transducers at 3 MHz. Journal of Athletic Training 42, 22-28.
Knight C.A., Rutledge C.R., Cox M.E., Acosta M., Hall S.J. (2001) Effect of superficial heat, deep heat, and active exercise warm-up on the extensibility of the plantar flexors. Physical Therapy 81, 1206-1214.
Kramer J.F. (1984) Ultrasound: Evaluation of its mechanical and thermal effects. Archives of Physical Medicine and Rehabilitation 65, 223-227.
Leung M.C.P., Ng G.Y.F., Yip K.K. (2006) Therapeutic ultrasound enhances medial collateral ligament repair in rats. Ultrasound in Medicine & Biology 32, 449-452.
McCutchan E.D., Demchak T.J., Brucker J.B. (2007) A comparison of the heating efficacy of the Autosound™ with traditional ultrasound methods. Journal of Athletic Training 42, S41-.
Merrick M.A., Bernard K.D., Devor S.T., Williams J.M. (2003) Identical 3-MHz ultrasound treatments with different devices produce different intramuscular temperatures: current treatment parameters may not be adequate. Journal of Orthopaedic Sports Physical Therapy 33, 379-385.
Myrer J.W., Measom G.J., Fellingham G.W. (2001) Intramuscular temperature rises with topical analgesics used as coupling agents during therapeutic ultrasound. Journal of Athletic Training 36, 20-26.
Noble J.G., Lee V., Griffeth-Noble F. (2007) Therapeutic ultrasound: the effects upon cutaneous blood flow in humans. Ultrasound in Medicine & Biology 33, 279-285.
Oates D., Draper D.O. (2006) Restoring wrist range of motion using ultrasound and mobilization: A case study. Athletic Therapy Today 11, 45-47.
Oshikoya C.A., Shultz S.J., Mistry D., Perrin D.H., Arnold B.L., Gansneder B.M. (2000) Effect of coupling medium temperature on rate of intramuscular temperature rise using continuous ultrasound. Journal of Athletic Training 35, 417-421.
Prentice W.E. (2003) Therapeutic modalities for sports medicine and athletic training. New York, NY. McGraw Hill.
Raso V.V.M., Barbieri C.H., Mazzer N., Fasan V.S. (2005) Can therapeutic ultrasound influence the regeneration of peripheral nerves?. Journal of Neuroscience Methods 142, 185-192.
Rose S., Draper D.O., Schulthies S.S., Durrant E. (1996) The stretching window part two: Rate of thermal decay in deep muscle following 1-MHz ultrasound. Journal of Athletic Training 31, 139-143.
Stein H., Lerner A. (2005) How does pulsed low-intensity ultrasound enhance fracture healing?. Orthopedics 28, 1161-1164.
Warden S.J. (2003) A new direction for ultrasound therapy in sports medicine. Sports Medicine 33, 95-107.
Warden S.J., Avin K.G., Beck E.M., DeWolf M.E., Hagemeier M.A., Martin K.M. (2006) Low-intensity pulsed ultrasound accelerates and a nonsteroidal anti-inflammatory drug delays knee ligament healing. American Journal of Sports Medicine 34, 1094-1102.
Weaver S.L., Demchak T.J., Stone M.B., Brucker J.B., Burr P.O. (2006) Effect of transducer velocity on intramuscular temperature during a 1-MHz ultrasound treatment. Journal of Orthopaedic Sports Physical Therapy 36, 320-325.








Back
|
PDF
|
Share