Table 1. Peak temperatures in the skin surface and muscle induced by microwave hyperthermia at two radio frequencies. Because a warm water bolus was placed on the thigh before treatment, the baseline skin temperature increased in the 434-MHz system.
Baseline T
°C
Peak T
°C
Temperature rise
°C
Time to peak T
min
434-MHz system
Skin surface
Mean (SD) 34.2 (1.4) * 39.2 (.5) 5.0 (1.5) * 10.7 (1.6)
Range 31.6-35.9 38.2-39.8 3.1-8.2 8.4-13. 7
Muscle
Mean (SD) 35.0 (1.3) 43.7 (.8) ** 8.9 (1.4) ** 9.9 (1.4) **
Range 33.2-37.9 42.8-45.4 6.3-11.4 8.4-12.5
2450-MHz system
Skin surface
Mean (SD) 32.3 (1.1) 39.2 (.8) 6.9 (1.5) 15.9 (7.3)
Range 30.8-33.9 38.0-40.6 4.1-8.3 4.8-30.2
Muscle
Mean (SD) 35.0 (.8) 41.1 (1.3) 6.1 (1.4) 20.0 (7.5)
Range 33.6-36.0 39.6-43.2 3.9-8.5 9.5-30.3
* and ** denote p < 0.05 and 0.01, respectively compared with 2450-MHz system.
* and ** denote p < 0.05 and 0.01, respectively compared with 2450-MHz system.
* and ** denote p < 0.05 and 0.01, respectively compared with 2450-MHz system.